DocumentCode :
235964
Title :
Relative humidity effects on rectangular waveguide thermistor sensors
Author :
Crowley, T.P.
Author_Institution :
Nat. Inst. of Stand. & Technol., Boulder, CO, USA
fYear :
2014
fDate :
24-29 Aug. 2014
Firstpage :
22
Lastpage :
23
Abstract :
A correlation between the relative humidity and the effective efficiency of several rectangular waveguide thermistor sensors has been observed. The difference between effective efficiency measurements made at 10 % and 30 % relative humidity can be 10 times the k = 1 reproducibility uncertainty of these devices. The time scale for the effective efficiency change is several days. The effect appears to be reversible. It has not been observed in any of NIST´s coaxial calorimeter standards.
Keywords :
humidity; measurement standards; power measurement; rectangular waveguides; sensors; thermistors; effective efficiency measurements; rectangular waveguide thermistor sensors; relative humidity effects; Humidity; Humidity measurement; Laboratories; Sensors; Standards; Thermistors; Uncertainty;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Precision Electromagnetic Measurements (CPEM 2014), 2014 Conference on
Conference_Location :
Rio de Janeiro
ISSN :
0589-1485
Print_ISBN :
978-1-4799-5205-2
Type :
conf
DOI :
10.1109/CPEM.2014.6898239
Filename :
6898239
Link To Document :