DocumentCode :
235984
Title :
Low-frequency characteristics of a thin-film multijunction thermal converter under vacuum
Author :
Amagai, Yasutaka ; Fujiki, Hiroyuki ; Shimizume, Koji ; Kishino, Katsumi ; Hidaka, Shigeru
Author_Institution :
Nat. Metrol. Inst. of Japan, Tsukuba, Japan
fYear :
2014
fDate :
24-29 Aug. 2014
Firstpage :
50
Lastpage :
51
Abstract :
We have evaluated the AC-DC transfer differences of our thin-film multijunction thermal converters with increased heat capacity under vacuum conditions to improve the low frequency characteristics. μμ The measured thermal time constant at about 0.001 Pa is increased up to 9 s from 3 s measured at ambient pressure. The AC-DC transfer differences evaluated by using a sampling method are less than 1 μV/V down to 10 Hz, and 10 μV/V down to 0.2 Hz, which is about ten times smaller compared with those obtained at ambient pressure.
Keywords :
power convertors; specific heat; thin film devices; AC-DC transfer differences; heat capacity; low-frequency characteristics; sampling method; thermal time constant; thin-film multijunction thermal converter; vacuum conditions; Frequency conversion; Frequency measurement; Heat transfer; Heating; Pressure measurement; Temperature measurement; Voltage measurement; AC-DC difference; Electrical variable measurement; thermal converters; thermal voltage converters;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Precision Electromagnetic Measurements (CPEM 2014), 2014 Conference on
Conference_Location :
Rio de Janeiro
ISSN :
0589-1485
Print_ISBN :
978-1-4799-5205-2
Type :
conf
DOI :
10.1109/CPEM.2014.6898253