Title :
Calibration of a thermal transfer standard using a pulse-driven Josephson voltage standard at frequencies up to 100 kHz
Author :
Hagen, T. ; Yi-hua Tang ; Lipe, Thomas E.
Author_Institution :
Nat. Meas. Inst., Lindfield, NSW, Australia
Abstract :
This paper describes the frequency extension from 10 kHz to 100 kHz of a system for the calibration of ac-dc thermal transfer standards against a pulse-driven Josephson voltage standard. Higher frequency operation requires careful consideration of the systematic error introduced by the on-chip inductance of the Josephson array, as well as careful selection of connecting cables. These considerations are discussed, and measurements using two different pulse sources are compared.
Keywords :
Josephson effect; calibration; inductance; measurement errors; measurement standards; voltage measurement; Josephson array; ac-dc thermal transfer standards; calibration; frequency 10 kHz to 100 kHz; on-chip inductance; pulse-driven Josephson voltage standard; systematic error; Arrays; Calibration; Coaxial cables; Generators; Radio frequency; Standards; Voltage measurement; Josephson arrays; ac-dc transfer; measurement techniques pulse driven Josephson junctions; thermal transfer standards;
Conference_Titel :
Precision Electromagnetic Measurements (CPEM 2014), 2014 Conference on
Conference_Location :
Rio de Janeiro
Print_ISBN :
978-1-4799-5205-2
DOI :
10.1109/CPEM.2014.6898255