Title :
Calibration of capacitance in the frequency range up to 1 MHz
Author :
Dan Bee Kim ; Hyung-Kew Lee ; Wan-Seop Kim
Author_Institution :
Korea Res. Inst. of Sci. & Technol., Daejeon, South Korea
Abstract :
This paper describes two different ways to calibrate the capacitance in the high frequency range up to 13 MHz. First, a dual source bridge was configured using two voltage sources and then the ratio of two capacitors was measured in the frequency range from 1 kHz to 1 MHz. The second method known as the Z-matrix method, allowed us to measure the capacitance value up to 13 MHz. The measured values were compared with the reference values calibrated by the PTB and Agilent Technologies, respectively, and they were also compared with each other. The cross-checks showed that both methods are reliable to an order of 10-5 at 1 MHz.
Keywords :
bridge circuits; calibration; capacitance measurement; capacitors; Agilent Technology; PTB; Z-matrix method; calibration; capacitance measurement; capacitor; dual source bridge; voltage source; Bridge circuits; Calibration; Capacitance; Capacitance measurement; Capacitors; Frequency measurement; Standards; Capacitance measurement; Z-matrix method; dual source bridge; high frequency;
Conference_Titel :
Precision Electromagnetic Measurements (CPEM 2014), 2014 Conference on
Conference_Location :
Rio de Janeiro
Print_ISBN :
978-1-4799-5205-2
DOI :
10.1109/CPEM.2014.6898273