Title :
The effect of humidity on the calibration of air-type capacitance standards
Author :
Hyung-Kew Lee ; Dan Bee Kim ; Wan-Seop Kim
Author_Institution :
Korea Res. Inst. of Sci. & Technol., Daejeon, South Korea
Abstract :
This paper describes our observation of the effect of humidity on a four-terminal-pair (4TP) air capacitance standard set. A standard set (Agilent 16380A) has been monitored four fourteen days at 1 kHz frequency using an AH 2500A capacitance bridge. Consistent increment of measured capacitance was observed for all the capacitors in the set as the humidity increases. In addition, the analysis on the humidity-dependent capacitance reveals that the capacitance value follows the humidity changes with a delay of about 0.7 day.
Keywords :
calibration; capacitance measurement; humidity; measurement standards; calibration; capacitance bridge; capacitance measurement; current 2500 A; four terminal pair air capacitance standard; frequency 1 kHz; humidity effect; Capacitance; Capacitance measurement; Capacitors; Humidity; Humidity measurement; Monitoring; Standards; Capacitance measurement; effect of humidity;
Conference_Titel :
Precision Electromagnetic Measurements (CPEM 2014), 2014 Conference on
Conference_Location :
Rio de Janeiro
Print_ISBN :
978-1-4799-5205-2
DOI :
10.1109/CPEM.2014.6898274