• DocumentCode
    2361033
  • Title

    Designing an alarm system for the stratification of risk of cardiac arrhythmias

  • Author

    Álvarez, E. ; Jiménez, J. ; Moleiro, F. ; Rodriguez, A.

  • Author_Institution
    Lab. de Fenomenos No-Lineales, UCV
  • fYear
    2008
  • fDate
    14-17 Sept. 2008
  • Firstpage
    1037
  • Lastpage
    1040
  • Abstract
    The idea of this research is to propose an alarm system that can alert the early establishment of an event paroxysmal ventricular tarquicardia (PVT), from study of heart rate variability (HRV). The study of the variability is support in the frequency of abnormal patterns prior to the events. We select and analyze in detail 24 hours records from dynamic electrocardiography (Holter). We selected 88 one-hour segments prior to the events of PVT, and once analyzed the system found an efficiency of 77% in early detection of the establishment of the arrhythmia, with an average prediction time of 43 plusmn 8 minutes distant from events, statistically significant result. To assess the percentage of false alarms we evaluate 103 hours from control healthy subjects, resulting in a specificity of 85%.
  • Keywords
    alarm systems; diseases; electrocardiography; medical signal detection; alarm system; cardiac arrhythmia; dynamic electrocardiography; heart rate variability; paroxysmal ventricular tarquicardia; risk stratification; time 103 hour; time 24 hour; Alarm systems; Cardiac disease; Cardiology; Control systems; Electrocardiography; Frequency; Heart rate variability; Laboratories; Medical control systems; System testing;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Computers in Cardiology, 2008
  • Conference_Location
    Bologna
  • ISSN
    0276-6547
  • Print_ISBN
    978-1-4244-3706-1
  • Type

    conf

  • DOI
    10.1109/CIC.2008.4749222
  • Filename
    4749222