DocumentCode :
236177
Title :
AC characterizations of current shunts
Author :
Kon, S. ; Yamada, Tomoaki
Author_Institution :
Nat. Metrol. Inst. of Japan, AIST, Tsukuba, Japan
fYear :
2014
fDate :
24-29 Aug. 2014
Firstpage :
244
Lastpage :
245
Abstract :
The dependence of cage-type AC shunt resistance and phase angle on frequency and input current is examined in this paper, for shunts with different current ratings. The AC resistance of shunts remains constant for up to 1 kHz, but the phase angle changes proportionally to the frequency. The results of current dependencies on AC resistance are consistent with each other within measurement uncertainties between 1 A and 10 A. The phase angle does not depend on the current, for values up to 10 A. The effects of current heating of shunts are also evaluated. The thermal time constant of shunts is found to be approximately 200 s and does not depend on the shunts and the input current frequency.
Keywords :
calibration; electric current measurement; electric resistance measurement; measurement uncertainty; cage-type AC shunt resistance; calibration; current heating; measurement uncertainty; phase angle; thermal time constant; Calibration; Current measurement; Current transformers; Measurement uncertainty; Resistance; Resistors; Standards; AC characteristic; calibration; current; phase angle; shunt resistor;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Precision Electromagnetic Measurements (CPEM 2014), 2014 Conference on
Conference_Location :
Rio de Janeiro
ISSN :
0589-1485
Print_ISBN :
978-1-4799-5205-2
Type :
conf
DOI :
10.1109/CPEM.2014.6898350
Filename :
6898350
Link To Document :
بازگشت