DocumentCode :
236187
Title :
10 V programmable Josephson voltage standard and its application in direct comparison with conventional Josephson voltage standard
Author :
Yi-hua Tang ; Sims, June E.
Author_Institution :
Nat. Inst. of Stand. & Technol., Gaithersburg, MD, USA
fYear :
2014
fDate :
24-29 Aug. 2014
Firstpage :
256
Lastpage :
257
Abstract :
This paper briefly describes the working principle of the 10 V programmable Josephson voltage standard (PJVS) that was developed at the National Institute of Standards and Technology (NIST) and how to use it in a direct comparison with a conventional Josephson voltage standard (JVS). Manual and automatic comparison methods were investigated to verify the agreement between the two types of Josephson standards. The 10 V PJVS will be used as a transfer standard in the 2014 JVS Interlaboratory Comparison (ILC) that is organized by the National Conference of Standard Laboratories International (NCSLI).
Keywords :
Josephson effect; measurement standards; voltage measurement; automatic comparison; manual comparison; programmable Josephson voltage standard; transfer standard; voltage 10 V; 1f noise; Floors; NIST; Protocols; Uncertainty; Voltage measurement; automatic comparison; conventional Josephson voltage standard; programmable Josephson voltage standard; uncertainty;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Precision Electromagnetic Measurements (CPEM 2014), 2014 Conference on
Conference_Location :
Rio de Janeiro
ISSN :
0589-1485
Print_ISBN :
978-1-4799-5205-2
Type :
conf
DOI :
10.1109/CPEM.2014.6898356
Filename :
6898356
Link To Document :
بازگشت