Abstract :
Thermal converters show significant ac-dc transfer differences at low frequencies due to nonlinearities of the thermocouples. It is assumed, that the ac-dc transfer differences at low frequencies are proportional to the input power. This assumption was proofed using an independent method with sampling techniques. A novel approach based on sinewave fitting is used to calculate the effective value oft the sampled signal from the samples. It takes into account the special situation of a metrological environment.
Keywords :
curve fitting; signal sampling; temperature measurement; thermocouples; waveform generators; AC-DC transfer difference; metrological environment; signal sampling; sinewave fitting; thermal converter; thermocouple nonlinearities; Approximation algorithms; Fitting; Frequency conversion; Frequency measurement; Uncertainty; Voltage measurement; Voltmeters; ac-dc transfer; curve fitting; signal sampling; thermal converter;