• DocumentCode
    2362234
  • Title

    Evaluation of the V(x) temperature stability in time-resolved ultrasonic measurements

  • Author

    Titov, S.A. ; Maev, R.G. ; Bogatchenkov, A.N.

  • Author_Institution
    Centre for Imaging Res. & Adv. Material Characterization, Windsor Univ., Ont., Canada
  • Volume
    1
  • fYear
    2001
  • fDate
    2001
  • Firstpage
    497
  • Abstract
    Thermal errors of time-resolved ultrasonic material characterization systems are considered. For V(z) data acquisition configuration, theoretically estimated the thermal error coefficient of leaky wave velocity was found to be 10-3 1/°C. It was shown that the reading of the sound velocity in immersion liquid is not required for V(x) data processing. The thermal error coefficient estimated experimentally in 20-50°C range with the developed V(x) system is 5·10-5 1/°C for a water-fused quartz interface
  • Keywords
    acoustic wave velocity; quartz; surface acoustic waves; thermal stability; water; 20 to 50 degC; SiO2-H2O; V(x) temperature stability; leaky wave velocity; sound velocity; thermal error coefficient; time-resolved ultrasonic material characterization; time-resolved ultrasonic measurements; water-fused quartz interface; Biological materials; Data acquisition; Equations; Frequency; Measurement errors; Stability; Temperature; Transducers; Ultrasonic variables measurement; Velocity measurement;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Ultrasonics Symposium, 2001 IEEE
  • Conference_Location
    Atlanta, GA
  • Print_ISBN
    0-7803-7177-1
  • Type

    conf

  • DOI
    10.1109/ULTSYM.2001.991671
  • Filename
    991671