DocumentCode
2362694
Title
Parameter extraction for diagnosing patent ductus arteriosus in preterm neonates using phonocardiography
Author
Balogh, Ádám T. ; Kovács, Ferenc
Author_Institution
Fac. of Inf. Technol., Pazmany Peter Catholic Univ., Budapest, Hungary
fYear
2010
fDate
7-10 Nov. 2010
Firstpage
1
Lastpage
2
Abstract
In this paper a method is described for parameter extraction from murmur related to patent ductus arteriosus (PDA) in case of preterm infants using phonocardiography. The relationship between the extracted time and frequency domain parameters and the medical parameters of the patent ductus arteriosus has been investigated. This analysis revealed that the frequency parameters of the murmur could be related to the diameter of the PDA and to the maximal blood flow velocity through the PDA. Nevertheless the results are promising, further work is needed for more reliable parameter estimation for robust verification of the mentioned relationships.
Keywords
acoustic signal processing; blood vessels; diseases; feature extraction; haemodynamics; medical signal processing; parameter estimation; phonocardiography; PDA diameter; PDA related murmur; frequency domain parameters; maximal blood flow velocity; parameter extraction; patent ductus arteriosus diagnosis; phonocardiography; preterm neonates; time domain parameters;
fLanguage
English
Publisher
ieee
Conference_Titel
Applied Sciences in Biomedical and Communication Technologies (ISABEL), 2010 3rd International Symposium on
Conference_Location
Rome
Print_ISBN
978-1-4244-8131-6
Type
conf
DOI
10.1109/ISABEL.2010.5702875
Filename
5702875
Link To Document