DocumentCode
236388
Title
Mutual inductance measurement with programmable Josephson system
Author
Yuan Gao ; Honghui Li ; Zengmin Wang
Author_Institution
Nat. Inst. of Metrol., Beijing, China
fYear
2014
fDate
24-29 Aug. 2014
Firstpage
460
Lastpage
461
Abstract
We developed a magnetic flux measurement system which is based on the binary Josephson system two years ago. After that, we continue researching on the improvement of magnetic flux measurement and start to focus on the precision measurement of the mutual inductance with the flux measurement system we built before. This paper introduces the mutual inductance measurement method which is based on the programmable Josephson system at NIM. Experimental results show that the type A uncertainty of the measurement of mutual inductances of 341 mH and 577 mH is less than 2 ppm, which demonstrates the feasibility and reliability of the flux and mutual inductance measurement method we proposed.
Keywords
inductance measurement; magnetic field measurement; measurement uncertainty; reliability; superconducting junction devices; NIM; binary Josephson system; magnetic flux measurement system; mutual inductance measurement uncertainty; precision measurement; programmable Josephson system; reliability; Coils; Current measurement; Inductance; Inductance measurement; Magnetic flux; Measurement uncertainty; Voltage measurement; Flux Measurement; Josephson voltage standard; Mutual Inductance;
fLanguage
English
Publisher
ieee
Conference_Titel
Precision Electromagnetic Measurements (CPEM 2014), 2014 Conference on
Conference_Location
Rio de Janeiro
ISSN
0589-1485
Print_ISBN
978-1-4799-5205-2
Type
conf
DOI
10.1109/CPEM.2014.6898458
Filename
6898458