• DocumentCode
    236424
  • Title

    Using digital multimeters in place of analog null meters for metrological applications

  • Author

    Faulkner, Neil ; Gust, Jeff

  • Author_Institution
    Fluke Calibration, Everett, WA, USA
  • fYear
    2014
  • fDate
    24-29 Aug. 2014
  • Firstpage
    496
  • Lastpage
    497
  • Abstract
    Historically, analog null meters have been used to determine zero voltage conditions in metrological applications. When digital multimeters (DMMs) became available with resolution comparable to null meters, they were adopted quickly, and analog null meters were set aside. Depending on the circuit and the DMM, significant measurement errors can be introduced. This paper discusses in detail investigations associated with one particular model of precision DMM when used as a null detector and provides measurement configuration details to minimize errors so that the DMM can out-perform analog null meters.
  • Keywords
    digital multimeters; measurement errors; minimisation; voltage measurement; DMM; analog null meters; digital multimeters; measurement error minimization; metrological applications; zero voltage conditions determination; Bridge circuits; Calibration; Detectors; Noise; Resistance; Standards; Voltage measurement; measurement techniques; null detectors; null meters; precision measurements;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements (CPEM 2014), 2014 Conference on
  • Conference_Location
    Rio de Janeiro
  • ISSN
    0589-1485
  • Print_ISBN
    978-1-4799-5205-2
  • Type

    conf

  • DOI
    10.1109/CPEM.2014.6898476
  • Filename
    6898476