• DocumentCode
    236434
  • Title

    A high frequency voltage standard from 30 kHz to 30 MHz based on thermal voltage converters

  • Author

    He Zhao ; Tian Wei ; Huang Jianming ; Zhang Yichi ; Guo Xiaotao

  • Author_Institution
    Nat. Inst. of Metrol., Beijing, China
  • fYear
    2014
  • fDate
    24-29 Aug. 2014
  • Firstpage
    506
  • Lastpage
    507
  • Abstract
    A high frequency voltage standard of NIM from 30 kHz to 30 MHz is designed based on thermal voltage converters. In this standard, seven TVCs of different sizes are used to cover the voltage range from 0.1 V to a maximum of 100 V with UHF vacuum thermocouples and non-inductive resistors. The new high-frequency standard has been shown to have an accuracy of 0.01% at 1 MHz and 1 V by comparing it with the NIM low-frequency standard. Comparisons with a commercial TVC suggest the new standard has an accuracy of 0.05% at 10 MHz and 0.1% at 30 MHz.
  • Keywords
    UHF detectors; UHF measurement; convertors; frequency standards; resistors; thermocouples; voltage measurement; NIM; TVC; UHF vacuum thermocouple; frequency 30 kHz to 30 MHz; high frequency voltage standard; low-frequency standard; noninductive resistor; thermal voltage converter; voltage 0.1 V to 100 V; Accuracy; Frequency conversion; Indexes; Metrology; Resistors; Standards; Uncertainty; AC-DC transfer standard; Thermal Voltage Converter; Voltage Standard;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements (CPEM 2014), 2014 Conference on
  • Conference_Location
    Rio de Janeiro
  • ISSN
    0589-1485
  • Print_ISBN
    978-1-4799-5205-2
  • Type

    conf

  • DOI
    10.1109/CPEM.2014.6898481
  • Filename
    6898481