Title :
The reference step method for DC voltage at 1 mV and 10 mV
Author :
Andersson, Bengt-Olof ; Early, Murray ; Eklund, Gunnar ; Gunnarsson, Ove ; Rydler, Karl-Erik
Author_Institution :
SP Tech. Res. Inst. of Sweden, Boras, Sweden
Abstract :
We have investigated the performance of a method called the Reference Step Method at low voltage levels. The method has been evaluated earlier at voltage levels from 0.1 V to 1 kV with excellent results. The advantages of this method are that it uses commonly available equipment and can be fully automated. We have made comparisons with three other low-voltage methods and the differences show that the uncertainties can be less than 30 μV/V at 1 mV and 3 μV/V at 10 mV, making it possible to use this method to perform automated calibration of calibrators with sufficient accuracy for calibration of high-end instruments.
Keywords :
calibration; measurement uncertainty; voltage measurement; automated calibration; low voltage measurement; measurement uncertainty; reference step method; voltage 0.1 V to 1 kV; voltage 1 mV; voltage 10 mV; Calibration; Gain measurement; Instruments; Low voltage; Standards; Uncertainty; DC voltage ratio; dc range calibrators; dc voltage build-up; dc voltage scale; low voltage measurements;
Conference_Titel :
Precision Electromagnetic Measurements (CPEM 2014), 2014 Conference on
Conference_Location :
Rio de Janeiro
Print_ISBN :
978-1-4799-5205-2
DOI :
10.1109/CPEM.2014.6898488