DocumentCode :
236551
Title :
Investigating linearity of a high voltage capacitance bridge
Author :
Suomalainen, E.-P. ; Hallstrom, Jari
Author_Institution :
MIKES, Espoo, Finland
fYear :
2014
fDate :
24-29 Aug. 2014
Firstpage :
620
Lastpage :
621
Abstract :
High voltage capacitance bridge can be calibrated at fixed ratios using traceable standard capacitors. Typical ratios to be calibrated are 10 to 1 and 100 to 1. Substitution method can be used to reduce bridge related errors and uncertainties. When a capacitor under calibration deviates significantly from the reference capacitor, substitution method is not any more applicable and linearity of a capacitance bridge has to be considered. This paper describes a method to investigate linearity of a high voltage capacitance bridge. The method is based on two reference capacitors and two simultaneously sampling voltmeters. Results of the linearity test of four largest decades of a current comparator capacitance bridge are presented.
Keywords :
bridge circuits; calibration; capacitors; measurement standards; voltage measurement; voltmeters; calibration; high voltage capacitance bridge; linearity test; reference capacitor; substitution method; traceable standard capacitors; voltmeters; Bridge circuits; Capacitance; Capacitors; Current measurement; Linearity; Uncertainty; Voltage measurement; Capacitance bridge; capacitance; high voltage; high-voltage techniques; sampling techniques;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Precision Electromagnetic Measurements (CPEM 2014), 2014 Conference on
Conference_Location :
Rio de Janeiro
ISSN :
0589-1485
Print_ISBN :
978-1-4799-5205-2
Type :
conf
DOI :
10.1109/CPEM.2014.6898538