Title :
Accuracy of pacing system analyzers for the determination of sinus rhythm amplitude in an implantable cardioverter-defibrillator
Author :
Cao, J. ; Whitman, TA ; Perttu, JS ; Huntwork, DE
Author_Institution :
Angeion Corp., Minneapolis, MN, USA
Abstract :
The objective of this study was to compare the beat-by-beat differences in R-wave amplitudes measured manually with two pacing system analyzers (PSAs) and with an implantable cardioverter-defibrillator (ICD). Intracardiac electrograms (EGMs) were obtained during normal sinus rhythm from 40 patients who experienced a range of ventricular and/or atrial arrhythmias. An in-vitro testing system was used to reconstruct these EGMs with uniform R-wave amplitudes and supply the output signals to a PSA, ICD or strip chart recorder. R-wave amplitudes measured by each method were highly correlated to the manual measurement (r>0.85). The average amplitude measured by the manual method was 7.0+/-3.1 mV. Amplitudes for the Angeion ICD, SeaMED PSA, and Pace Medical PSA were 6.9+/-3.0, 8.3+/-3.7, 5.1+/-2.3 mV, respectively. A minimum of 8 mV measured with a PSA will most likely guarantee a minimum of 5 mV amplitude measured manually
Keywords :
defibrillators; electrocardiography; medical signal processing; pacemakers; prosthetics; signal reconstruction; -3.7 to 8.3 mV; R-wave amplitudes; atrial arrhythmias; beat-by-beat differences; implantable cardioverter-defibrillator; in-vitro testing system; intracardiac electrograms; output signals; pacing system analyzers accuracy; strip chart recorder; uniform R-wave amplitudes; ventricular arrhythmias; Biomedical measurements; Cardiology; Extraterrestrial measurements; Fibrillation; Filters; In vitro; Instruments; Rhythm; Strips; System testing;
Conference_Titel :
Computers in Cardiology 1998
Conference_Location :
Cleveland, OH
Print_ISBN :
0-7803-5200-9
DOI :
10.1109/CIC.1998.731787