Title :
Accurate measurement of AC voltage in audio band using Agilent 3458A sampling capability
Author :
Lapuh, Rado ; Voljc, Bostjan ; Lindic, Matjaz
Author_Institution :
SIQ - Slovenian Inst. of Quality & Metrol., Ljubljana, Slovenia
Abstract :
The paper describes frequency dependence correction of the Agilent 3458A DMM in DCV sampling mode. The biggest error source identified is the input filter roll-off, which was measured and evaluated with an asynchronous sampling technique using phase sensitive sine fit algorithm. Corrected response is flat within 50 μV/V in the whole audio frequency band.
Keywords :
digital multimeters; electric current measurement; sampling methods; voltage measurement; AC voltage measurement; Agilent 3458A DMM; Agilent 3458A sampling capability; DCV sampling mode; asynchronous sampling technique; audio frequency band; error source; frequency dependence correction; input filter roll-off; phase sensitive sine fit algorithm; Apertures; Bandwidth; Frequency conversion; Frequency dependence; Frequency measurement; Uncertainty; Voltage measurement; Measurement of ac voltage; PSFE algorithm; asynchronous sampling; audio band; calibration; correction; frequency dependence; nput filter roll-off; uncertainty;
Conference_Titel :
Precision Electromagnetic Measurements (CPEM 2014), 2014 Conference on
Conference_Location :
Rio de Janeiro
Print_ISBN :
978-1-4799-5205-2
DOI :
10.1109/CPEM.2014.6898598