  • Abstract
    This paper presents an automated microvolt Josephson voltage set-up developed at NIM. It is based on two programmable Josephson SINIS arrays, and can be used to calibrate the lower ranges of nanovoltmeters for low level DC measurements. A low noise bias source developed at NIM has been used and the step flatness of the arrays was verified carefully. The preliminary results of the measurements set at 3 μV show that the difference between the measured values and the theoretical value is 0.03 nV with the uncertainty (k = 1) of 0.10 nV.
  • Keywords
    Josephson effect; calibration; electric current measurement; measurement uncertainty; surface topography measurement; voltage measurement; voltmeters; NIM; automated microvolt Josephson voltage set-up; calibration; low level DC measurement; low noise bias source; measurement uncertainty; nanovoltmeter; programmable Josephson SINIS array; step flatness; voltage 0.03 nV; voltage 0.10 nV; Calibration; Current measurement; Josephson junctions; Junctions; Measurement uncertainty; Uncertainty; Voltage measurement; DC voltage calibration; Programmable Josephson array; low level measurement; microvolt;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements (CPEM 2014), 2014 Conference on
  • Conference_Location
    Rio de Janeiro
  • ISSN
    0589-1485
  • Print_ISBN
    978-1-4799-5205-2
  • Type

    conf

  • DOI
    10.1109/CPEM.2014.6898600
  • Filename
    6898600