• DocumentCode
    2366544
  • Title

    Interconnects for device ESD protection

  • Author

    Jarrett, Tim ; Unger, Burt

  • Author_Institution
    Cardiac Pacemakers Inc., St. Paul, MN, USA
  • fYear
    1996
  • fDate
    10-12 Sept. 1996
  • Firstpage
    123
  • Lastpage
    129
  • Abstract
    A conductive spring loaded shorting bar on the leads of a TO-220 MOSFET was found to protect it from ESD failures. The MOSFET is used in an implantable cardioverter defibrillator device where quality and reliability is essential. Pre-production samples revealed significant ESD failures, thought to be associated with a mechanical shearing operation. Analysis suggests CDM failures, since testing revealed a low CDM threshold. A custom designed removable shorting interconnect raised all ESD thresholds by at least an order of magnitude.
  • Keywords
    MOSFET; defibrillators; electrostatic discharge; protection; CDM testing; ESD failure; TO-220 MOSFET; conductive spring loaded shorting bar; implantable cardioverter defibrillator; interconnect; mechanical shearing; protection; Defibrillators; Electrostatic discharges; Implantable biomedical devices; Interconnections; MOSFETs; Protection;