• DocumentCode
    236665
  • Title

    Highly accurate differential AC voltage measurements with a single DC voltage reference

  • Author

    Kurten Ihlenfeld, Waldemar G. ; Landim, Regis P.

  • Author_Institution
    Inst. Nac. de Metrol., Qualidade e Tecnol. (INMETRO), Rio de Janeiro, Brazil
  • fYear
    2014
  • fDate
    24-29 Aug. 2014
  • Firstpage
    766
  • Lastpage
    767
  • Abstract
    This paper describes a measurement setup employed to investigate differential voltage measurements to be done with a spectrally pure sine wave and a dc reference. The system enables adjustment, to a very high accuracy, the start phase of ac sampling. Repeatabilities of some 0.1 μV/V were obtained with the help of adaptive digital control.
  • Keywords
    adaptive control; digital control; electric current measurement; reference circuits; voltage measurement; AC sampling; DC reference; adaptive digital control; differential AC voltage measurement; sine wave reference; single DC voltage reference; Band-pass filters; Capacitors; Distortion measurement; Harmonic analysis; Power harmonic filters; Synchronization; Voltage measurement; Ac voltage; Josephson; chopped dc reference; digital signal processing; measurement standards; sampling technique;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements (CPEM 2014), 2014 Conference on
  • Conference_Location
    Rio de Janeiro
  • ISSN
    0589-1485
  • Print_ISBN
    978-1-4799-5205-2
  • Type

    conf

  • DOI
    10.1109/CPEM.2014.6898611
  • Filename
    6898611