DocumentCode
2367553
Title
Multi-frequency interstitial ultrasound applicator for conformal thermal therapy
Author
Chopra, Rajiv ; Luginbuhl, Chris ; Weymouth, Alfred J. ; Foster, F. Stuart ; Bronskill, Michael J.
Author_Institution
Dept. of Med. Biophys., Toronto Univ., Ont., Canada
Volume
2
fYear
2001
fDate
2001
Firstpage
1335
Abstract
The ability to deliver conformal thermal therapy was investigated for multi-frequency interstitial heating applicators incorporating planar transducers and device rotation. Thermal lesions were generated in excised tissue and the influence of exposure parameters such as frequency, power, and device rotation was studied. Lesion radius was dependent on all exposure parameters studied, and a maximum radius of 18.0±0.6 mm was achieved. This applicator design and delivery strategy make conformal interstitial thermal therapy possible
Keywords
biomedical ultrasonics; biothermics; radiation therapy; tumours; ultrasonic transducers; 17.4 to 18.6 mm; conformal thermal therapy; device rotation; exposure parameters; lesion radius; multi-frequency interstitial ultrasound applicator; planar transducers; thermal lesions; tumours; Acoustic transducers; Applicators; Force measurement; Frequency; Heating; Lesions; Medical treatment; Power generation; Ultrasonic imaging; Ultrasonic transducers;
fLanguage
English
Publisher
ieee
Conference_Titel
Ultrasonics Symposium, 2001 IEEE
Conference_Location
Atlanta, GA
Print_ISBN
0-7803-7177-1
Type
conf
DOI
10.1109/ULTSYM.2001.991967
Filename
991967
Link To Document