Abstract :
A new real-time high frequency ultrasound scanner has been developed to easily image superficial structures and estimate blood flow, with several major improvements. A great limitation of previous high frequency systems has been the requirement to water-couple the mechanically-scanned transducer to the object to be imaged, and slowly scan the transducer using a fixed motion system. In this system, we have integrated a high-speed actuator used in conjunction with a low mass transducer enclosed in a hand-held scanhead, which allows direct placement on the eyelid or skin. The design of this system is optimized for high frame rates and clinical versatility. Implemented on this system are B-scan, M-mode, and swept-scan imaging modes, with velocity estimation in the latter two. A series of experiments were performed to characterize system performance. The system provides a frame rate of 10 fps, which is partially limited by the execution time of the software. Phantom images from the scanner have 8-bit resolution with a SNR in excess of 35 dB, providing sharp, high-resolution images
Keywords :
biomedical ultrasonics; blood flow measurement; image resolution; skin; ultrasonic transducers; 8 bit; B-scan modes; M-mode modes; SNR; blood flow; clinical versatility; eyelid; frame rate; frame rates; hand-held scanhead; high frequency ultrasound scanner; high-resolution images; high-speed actuator; low mass transducer; skin; superficial structures; swept-scan imaging modes; Actuators; Blood flow; Design optimization; Eyelids; Frequency estimation; High-resolution imaging; Skin; System performance; Ultrasonic imaging; Ultrasonic transducers;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Ultrasonics Symposium, 2001 IEEE
Conference_Location :
Atlanta, GA
Print_ISBN :
0-7803-7177-1
Type :
conf
DOI :
10.1109/ULTSYM.2001.991996
Filename :
991996
Link To Document :
بازگشت