Title :
Ultrasound second harmonic imaging with a weighted chirp signal
Author :
Kim, D.Y. ; Lee, J.C. ; Kwon, S.J. ; Song, T.K.
Author_Institution :
Dept. of Electron. Eng., Sogang Univ., Seoul, South Korea
Abstract :
In this paper, a new harmonic imaging technique is proposed and evaluated experimentally. In the proposed method, a weighted chirp signal with a Hanning window is transmitted. The RF samples obtained on each array element are individually compressed by correlating with the reference signal defined as the 2nd harmonic component ( 2f0) of a transmitted chirp signal generated in a square-law system. The proposed method uses the compressed 2f0 component to form an image, for which the crosscorrelation term should be suppressed below at least -50 dB. Another advantage of the proposed method is that the SNR of 2nd harmonic imaging can be improved without limitation by increasing the duration of the chirp signal
Keywords :
biomedical ultrasonics; chirp modulation; harmonics; medical image processing; ultrasonic imaging; Hanning window; SNR; US second harmonic imaging; array element; medical imaging; ultrasound imaging system; weighted chirp signal; Band pass filters; Chirp; Image coding; Power harmonic filters; Pulse compression methods; Pulse inverters; RF signals; Radio frequency; Signal generators; Ultrasonic imaging;
Conference_Titel :
Ultrasonics Symposium, 2001 IEEE
Conference_Location :
Atlanta, GA
Print_ISBN :
0-7803-7177-1
DOI :
10.1109/ULTSYM.2001.991999