    Preliminary results of a raster scanning beam delivery system

  • Author

    Renner, T.R. ; Chu, W. ; Ludewigt, B. ; Halliwell, J. ; Nyman, M. ; Singh, R.P. ; Stover, G.D. ; Stradtner, R.

  • Author_Institution
    Lawrence Berkeley Lab., California Univ., CA, USA
  • fYear
    1989
  • fDate
    20-23 Mar 1989
  • Firstpage
    672
  • Abstract
    A beam delivery system using a raster scanning technique has been developed to create large radiation fields with light ion beams for radiation treatment of cancer patients. Radiation fields up to 20 cm by 30 cm have been produced with a beam of Ne ions having a magnetic rigidity of 6.9 T-m. The radiation fields produced by this method appear to be less sensitive to beam position and beam shape than the wobbler dynamic beam delivery system, but require more stringent beam spill requirements to insure uniform radiation fields
  • Keywords
    beam handling equipment; beam handling techniques; radiation therapy; Ne ions; cancer patients; large radiation fields; light ion beams; magnetic rigidity; radiation fields; raster scanning beam delivery; stringent beam spill; wobbler dynamic beam; Computer displays; Control systems; Detectors; Frequency; Ion beams; Ionization chambers; Magnetic field measurement; Magnets; Power supplies; Pulse measurements;