• DocumentCode
    2374378
  • Title

    Computations of the accuracy of VSM

  • Author

    Shau-Wei Hsu

  • Author_Institution
    Center for Meas. Stand., Ind. Technol. Res. Inst., Hsinchu, Taiwan
  • fYear
    1996
  • fDate
    17-21 June 1996
  • Firstpage
    38
  • Lastpage
    39
  • Abstract
    The accuracy of the measurements by vibrating-sample magnetometer (VSM) has been studied by the computations of the sensitivity function. The arrangement of the coils is important to approach a wide spatial-uniformity of the sensitivity function. The errors due to the directional displacements of the sphere are lower than 2% if the mis-alignments are controlled to within some specific values. The uncertain radius of the sphere is also an origin of errors.
  • Keywords
    error analysis; magnetometers; measurement errors; performance evaluation; sensitivity analysis; VSM; accuracy; directional displacements; errors; misalignments; sensitivity function; spatial-uniformity; vibrating-sample magnetometer; Calibration; Coils; Displacement control; Error correction; Magnetic fields; Magnetic moments; Measurement standards; Open loop systems; SQUID magnetometers; Vibration measurement;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements Digest, 1996 Conference on
  • Conference_Location
    Braunschweig, Germany
  • Print_ISBN
    0-7803-3376-4
  • Type

    conf

  • DOI
    10.1109/CPEM.1996.546547
  • Filename
    546547