DocumentCode :
2376063
Title :
Swallowable or implantable body temperature telemeter-body temperature radio pill
Author :
Zhou, Guan-Xiong
Author_Institution :
Inst. of Environ. Health & Eng., Chinese Acad. of Preventative Med., Beijing, China
fYear :
1989
fDate :
27-28 Mar 1989
Firstpage :
165
Lastpage :
166
Abstract :
The author introduces a body-temperature-transmitter-body-temperature-radio pill which can be swallowed or implanted. It shows the internal temperature under natural conditions with an accuracy of ±0.1°C. The dimensions are 6.5 mm diameter and 26-28 mm length. Using a 5.5 mm diameter×3.0 mm thick lithium battery, the pill can operated continuously for one month
Keywords :
biomedical equipment; biothermics; telemetering equipment; temperature measurement; 8.33×10-2 yr; Li battery; body temperature radio pill; implantable body temperature telemeter; internal temperature; medical telemeter; swallowable device; Animals; Batteries; Circuits; Digestive system; Humans; Radio transmitters; Silicon; Temperature; Testing; Transducers;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Bioengineering Conference, 1989., Proceedings of the 1989 Fifteenth Annual Northeast