Factors affecting the accuracy of volume-oscillometric blood pressure measurement during partial pressurization of the wrist
Author
Kim, Jong Pal ; Kim, Youn Ho ; Bae, Sangkon ; Kim, Seok Chan ; Shin, Kunsoo
Author_Institution
Samsung Adv. Inst. of Technol., South Korea
fYear
2009
fDate
3-6 Sept. 2009
Firstpage
721
Lastpage
724
Abstract
We compared the volume-oscillometric responses of the airbag pressure sensor and the contact force sensor across and along the radial artery on the wrist during partial pressurization by an airbag. Because of the anatomic structure and non-uniform pressurization pressure distribution, elongated and shifted oscillometric pressure waveform envelope variations are observed. For the contact force sensors directly above the radial artery, S-shaped pressurization curves can be seen possibly due to temporal softening of the radial artery stiffness at near zero transmural pressure. These differences in the shape of oscillometric envelope as well as pressurization curve may be the leading factors for inaccuracies of volume-oscillometric blood pressure measurement by partial pressurization method using an airbag.
Keywords
blood pressure measurement; blood vessels; force sensors; pressure sensors; airbag pressure sensor; contact force sensor; partial pressurization; radial artery; stiffness; volume-oscillometric blood pressure measurement; wrist; Blood Pressure; Blood Pressure Determination; Blood Pressure Monitors; Computer-Aided Design; Equipment Design; Equipment Failure Analysis; Humans; Oscillometry; Pressure; Radial Artery; Reproducibility of Results; Sensitivity and Specificity;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 2009. EMBC 2009. Annual International Conference of the IEEE
Conference_Location
Minneapolis, MN
ISSN
1557-170X
Print_ISBN
978-1-4244-3296-7
Electronic_ISBN
1557-170X
Type
conf
DOI
10.1109/IEMBS.2009.5332659
Filename
5332659
Link To Document