• DocumentCode
    2378709
  • Title

    Electrode design for high frequency block: Effect of bipolar separation on block thresholds and the onset response

  • Author

    Ackermann, D. Michael, Jr. ; Foldes, Emily L. ; Bhadra, Niloy ; Kilgore, Kevin L.

  • Author_Institution
    Dept. of Biomed. Eng., Case Western Reserve Univ., Cleveland, OH, USA
  • fYear
    2009
  • fDate
    3-6 Sept. 2009
  • Firstpage
    654
  • Lastpage
    657
  • Abstract
    The delivery of high frequency alternating currents (HFAC) to peripheral nerves has been shown to produce a rapid and reversible nerve conduction block at the site of the electrode, and holds therapeutic promise for diseases associated with undesired or pathological neural activity. It has been known since 1939 that the configuration of an electrode used for nerve block can impact the quality of the block, but to date no formal study of the impact of electrode design on high frequency nerve block has been performed. Using a mammalian small animal model, it is demonstrated that the contact separation distance for a bipolar nerve cuff electrode can impact two important factors related to high frequency nerve block: the amplitude of HFAC required to block the nerve (block threshold), and the degree to which the transient "onset response" which always occurs when HFAC is first applied to peripheral nerves, is present. This study suggests that a bipolar electrode with a separation distance of 1.0 mm minimizes current delivery while producing high frequency block with a minimal onset response in the rat sciatic nerve.
  • Keywords
    bioelectric phenomena; biomedical electrodes; diseases; neurophysiology; patient treatment; bipolar nerve cuff electrode; bipolar separation; block thresholds; disease therapy; distance 1.0 mm; electrode design; high frequency alternating currents; nerve conduction block; onset response; pathological neural activity; peripheral nerve; sciatic nerve; Action Potentials; Animals; Differential Threshold; Electric Stimulation; Electrodes, Implanted; Electromagnetic Fields; Equipment Design; Equipment Failure Analysis; Nerve Block; Neural Conduction; Rats; Rats, Sprague-Dawley; Sciatic Nerve; Sensitivity and Specificity;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 2009. EMBC 2009. Annual International Conference of the IEEE
  • Conference_Location
    Minneapolis, MN
  • ISSN
    1557-170X
  • Print_ISBN
    978-1-4244-3296-7
  • Electronic_ISBN
    1557-170X
  • Type

    conf

  • DOI
    10.1109/IEMBS.2009.5332738
  • Filename
    5332738