• DocumentCode
    2382084
  • Title

    Cardiolock2: Parallel singularities for the design of an active heart stabilizer

  • Author

    Bachta, Wael ; Renaud, Pierre ; Laroche, Edouard ; Gangloff, Jacques

  • Author_Institution
    LSIIT, Univ. Louis Pasteur, Illkirch, France
  • fYear
    2009
  • fDate
    12-17 May 2009
  • Firstpage
    3839
  • Lastpage
    3844
  • Abstract
    In this paper, the design of a new active cardiac stabilizer, Cardiolock2, is presented. Following the proof of concept Cardiolock [7], this device allows an active stabilization of the surface of a beating heart in two directions, which is considered sufficient for a complete stabilization. Piezoelectric actuation is combined with a compliant architecture to obtain high dynamics and accuracy. A remote center of motion is obtained with a serial architecture, and parallel mechanisms in configurations close to singularity are used to increase the workspace. A kinematic analysis is first presented, before detailing the main properties of the device and the current development of the prototype.
  • Keywords
    medical control systems; piezoelectric actuators; surgery; Cardiolock2; active heart stabilizer; coronary artery bypass grafting; heart surgery; piezoelectric actuation; Arteries; Cardiology; Heart; Kinematics; Minimally invasive surgery; Prototypes; Robotics and automation; Surges; Surgical instruments; Yarn;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Robotics and Automation, 2009. ICRA '09. IEEE International Conference on
  • Conference_Location
    Kobe
  • ISSN
    1050-4729
  • Print_ISBN
    978-1-4244-2788-8
  • Electronic_ISBN
    1050-4729
  • Type

    conf

  • DOI
    10.1109/ROBOT.2009.5152483
  • Filename