• DocumentCode
    2387821
  • Title

    Fine Granule Evaluation Model from Coarse Granule in Medical Ultrasonic System

  • Author

    Hata, Yutaka ; Kondo, Katsuya ; Kobashi, Syoji

  • Author_Institution
    Univ. of Hyogo, Hyogo
  • fYear
    2007
  • fDate
    2-4 Nov. 2007
  • Firstpage
    617
  • Lastpage
    617
  • Abstract
    In this paper, we describe medical ultrasonic systems, which evaluate a fine granule from a coarse granule. In ultrasonic systems, a measurement limitation is known as an object larger than the half-wavelength. However, we need to measure the object beyond the limitation. First, we describe an ultrasonic system for identifying a surface roughness. A fine granule evaluation model is made form the experiment. An ultrasonic system developed from the model can evaluate the roughness of a human imitated phantom with 89% accuracy. Second, we describe an ultrasonic system for identifying kinetic viscosity of liquid. The fine granule evaluation model is derived from a wave equation theoretically. The experimental results show that the developed system from the model can evaluate the viscosity.
  • Keywords
    biomedical ultrasonics; surface roughness; viscosity; coarse granule; fine granule evaluation model; half-wavelength; human imitated phantom; liquid kinetic viscosity; measurement limitation; medical ultrasonic system; surface roughness; Biological system modeling; Biomedical imaging; Blood; Humans; Imaging phantoms; Medical diagnostic imaging; Rough surfaces; Surface roughness; Ultrasonic variables measurement; Viscosity;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Granular Computing, 2007. GRC 2007. IEEE International Conference on
  • Conference_Location
    Fremont, CA
  • Print_ISBN
    978-0-7695-3032-1
  • Type

    conf

  • DOI
    10.1109/GrC.2007.43
  • Filename