Polyimide-based multi-channel arrayed electrode for measuring EEG signal on the skull of mouse
Author
Baek, Dong-Hyun ; Lee, Eun-Joong ; Moon, Jin-Hee ; Choi, Jee Hyun ; Pak, James Jungho ; Lee, Sang-Hoon
Author_Institution
Dept. of Biomed. Eng., Korea Univ., Seoul, South Korea
fYear
2009
fDate
3-6 Sept. 2009
Firstpage
7022
Lastpage
7025
Abstract
In this paper, we have developed 40 channel multiple electrodes mounted on the surface of mouse´s skull using polyimide substrate and tested its performance by measuring EEG signals. The recording site of the electrode was electroplated by Pt to enhance both contact impedance and adhesive strength by applying proper current, cleaning surface and removing H2 gas bubbles. For in vivo test, the electrode was placed on the skull of F1 mouse and EEG signals were measured. We observed the suitability of electrode for measuring EEG signals from multiple areas on the skull. The spectrum of EEG signal to change was observed by urethane administration.
Keywords
adhesion; biomedical electrodes; electroencephalography; electroplating; polymers; EEG signal; adhesive strength; contact impedance; electroplating; mouse skull; multichannel arrayed electrode; polyimide; urethane; Animals; Biomedical Engineering; Electric Impedance; Electrodes; Electroencephalography; Equipment Design; Mice; Mice, Inbred C57BL; Platinum; Resins, Synthetic; Signal Processing, Computer-Assisted; Skull;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 2009. EMBC 2009. Annual International Conference of the IEEE
Conference_Location
Minneapolis, MN
ISSN
1557-170X
Print_ISBN
978-1-4244-3296-7
Electronic_ISBN
1557-170X
Type
conf
DOI
10.1109/IEMBS.2009.5333320
Filename
5333320
Link To Document