• DocumentCode
    2391353
  • Title

    New in-situ calibration of diode detectors used in six-port network analyzers

  • Author

    Colef, Gabriel ; Karmel, Paul R. ; Ettenberg, Morris

  • Author_Institution
    Dept. of Electr. Eng., City Coll. of New York, NY, USA
  • fYear
    1989
  • fDate
    25-27 Apr 1989
  • Firstpage
    533
  • Lastpage
    535
  • Abstract
    The authors present a novel method of calibrating the diode detectors used in six-port network analyzers, without disconnecting the diodes. The method uses only a precision variable attenuator or a power meter with no additional directional coupler. The important feature of this method is that no actual measurement of power is required to calibrate the diode detectors for six-port operation. It is very well suited to in-situ calibration of the diode detection circuits for operation of the six-port with pulse signals when the average power detected by the power meter drops to very low levels for low duty cycles. In these cases, the precision variable attenuator method is required. Measurement results are presented
  • Keywords
    attenuators; calibration; microwave detectors; network analysers; semiconductor diodes; diode detectors; in-situ calibration; power meter; precision variable attenuator; six-port network analyzers; Attenuation measurement; Attenuators; Calibration; Diodes; Directional couplers; Envelope detectors; Intelligent networks; Power measurement; Testing; Voltage;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Instrumentation and Measurement Technology Conference, 1989. IMTC-89. Conference Record., 6th IEEE
  • Conference_Location
    Washington, DC
  • Type

    conf

  • DOI
    10.1109/IMTC.1989.36917
  • Filename
    36917