Abstract :
The high-static-pressure metrological characteristics of 1-GPa manganin resistance gauges are investigated. Nonlinearity is measured by means of a secondary standard-free piston gauge. A second-order mathematical-model is established, with an interpolation accuracy reaching 0.07%, as verified by calibration tests. The results obtained indicate that manganin resistance gauges exhibit very good stability, repeatability, and reproducibility
Keywords :
calibration; copper alloys; high-pressure techniques; interpolation; manganese alloys; nickel alloys; pressure transducers; 1 GPa; Cu-Mn-Ni; calibration tests; high pressure techniques; interpolation accuracy; manganin resistance gauges; nonlinearity; repeatability; reproducibility; second-order mathematical-model; secondary standard-free piston gauge; stability; Calibration; Chemistry; Electrical resistance measurement; Interpolation; Pistons; Pressure measurement; Temperature distribution; Temperature sensors; Wire; Wounds;