• DocumentCode
    2392446
  • Title

    Establishment of math-model and analysis of interpolation accuracy on 1-GPa manganin resistance gauges

  • Author

    Tian-You, Yuan ; Yan-Chong, Lu ; Bao-Hui, Yuan

  • Author_Institution
    Xian Modern Chem. Inst., China
  • fYear
    1989
  • fDate
    25-27 Apr 1989
  • Firstpage
    556
  • Lastpage
    558
  • Abstract
    The high-static-pressure metrological characteristics of 1-GPa manganin resistance gauges are investigated. Nonlinearity is measured by means of a secondary standard-free piston gauge. A second-order mathematical-model is established, with an interpolation accuracy reaching 0.07%, as verified by calibration tests. The results obtained indicate that manganin resistance gauges exhibit very good stability, repeatability, and reproducibility
  • Keywords
    calibration; copper alloys; high-pressure techniques; interpolation; manganese alloys; nickel alloys; pressure transducers; 1 GPa; Cu-Mn-Ni; calibration tests; high pressure techniques; interpolation accuracy; manganin resistance gauges; nonlinearity; repeatability; reproducibility; second-order mathematical-model; secondary standard-free piston gauge; stability; Calibration; Chemistry; Electrical resistance measurement; Interpolation; Pistons; Pressure measurement; Temperature distribution; Temperature sensors; Wire; Wounds;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Instrumentation and Measurement Technology Conference, 1989. IMTC-89. Conference Record., 6th IEEE
  • Conference_Location
    Washington, DC
  • Type

    conf

  • DOI
    10.1109/IMTC.1989.36922
  • Filename
    36922