DocumentCode
2395352
Title
Combined direct current and high frequency nerve block for elimination of the onset response
Author
Kilgore, Kevin L. ; Foldes, Emily A. ; Ackermann, D. Michael ; Bhadra, Niloy
Author_Institution
Dept. of Biomed. Eng., Case Western Reserve Univ., Cleveland, OH, USA
fYear
2009
fDate
3-6 Sept. 2009
Firstpage
197
Lastpage
199
Abstract
Nerve conduction in peripheral mammalian nerves can be blocked by high frequency alternating current (HFAC) waveforms. However, one of the disadvantages of HFAC block is that it produces an intense burst of firing in the nerve when the HFAC is first turned on. This is a significant obstacle to the clinical implementation of HFAC block. In this paper we present a method to produce HFAC block without the onset response, using a combination of direct current (DC) and HFAC block. This method was experimentally evaluated in an in-vivo mammalian model. Successful no-onset HFAC block was obtained using a DC block of 200 muA and an HFAC block of 30 kHz at 10 Vp-p. This may allow HFAC block to be used in clinical applications for pain relief.
Keywords
bioelectric phenomena; neurophysiology; current 200 muA; direct current; frequency 30 kHz; high frequency alternating current; nerve conduction; peripheral mammalian nerves; Animals; Electric Stimulation; Electric Stimulation Therapy; Muscle Contraction; Muscle, Skeletal; Nerve Block; Rats; Rats, Sprague-Dawley; Sciatic Nerve;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 2009. EMBC 2009. Annual International Conference of the IEEE
Conference_Location
Minneapolis, MN
ISSN
1557-170X
Print_ISBN
978-1-4244-3296-7
Electronic_ISBN
1557-170X
Type
conf
DOI
10.1109/IEMBS.2009.5333642
Filename
5333642
Link To Document