• DocumentCode
    2395352
  • Title

    Combined direct current and high frequency nerve block for elimination of the onset response

  • Author

    Kilgore, Kevin L. ; Foldes, Emily A. ; Ackermann, D. Michael ; Bhadra, Niloy

  • Author_Institution
    Dept. of Biomed. Eng., Case Western Reserve Univ., Cleveland, OH, USA
  • fYear
    2009
  • fDate
    3-6 Sept. 2009
  • Firstpage
    197
  • Lastpage
    199
  • Abstract
    Nerve conduction in peripheral mammalian nerves can be blocked by high frequency alternating current (HFAC) waveforms. However, one of the disadvantages of HFAC block is that it produces an intense burst of firing in the nerve when the HFAC is first turned on. This is a significant obstacle to the clinical implementation of HFAC block. In this paper we present a method to produce HFAC block without the onset response, using a combination of direct current (DC) and HFAC block. This method was experimentally evaluated in an in-vivo mammalian model. Successful no-onset HFAC block was obtained using a DC block of 200 muA and an HFAC block of 30 kHz at 10 Vp-p. This may allow HFAC block to be used in clinical applications for pain relief.
  • Keywords
    bioelectric phenomena; neurophysiology; current 200 muA; direct current; frequency 30 kHz; high frequency alternating current; nerve conduction; peripheral mammalian nerves; Animals; Electric Stimulation; Electric Stimulation Therapy; Muscle Contraction; Muscle, Skeletal; Nerve Block; Rats; Rats, Sprague-Dawley; Sciatic Nerve;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 2009. EMBC 2009. Annual International Conference of the IEEE
  • Conference_Location
    Minneapolis, MN
  • ISSN
    1557-170X
  • Print_ISBN
    978-1-4244-3296-7
  • Electronic_ISBN
    1557-170X
  • Type

    conf

  • DOI
    10.1109/IEMBS.2009.5333642
  • Filename
    5333642