    A sensing tube with an integrated piezoelectric flow sensor for liver transplantation

  • Author

    Jung, Wooseok ; Li, Chunyan ; Kim, Dong-Sik ; Ahn, Chong H.

  • Author_Institution
    Dept. of Electr. & Comput. Eng., Univ. of Cincinnati, Cincinnati, OH, USA
  • fYear
    2009
  • fDate
    3-6 Sept. 2009
  • Firstpage
    4469
  • Lastpage
    4472
  • Abstract
    An innovative sensing tube with an integrated piezoelectric flow sensor to measure flow rate in portal vein during liver transplantation has been developed and characterized in this work. This smart tube aims to measure flow rates in the portal vein in a real-time continuous format with high reliability. The PVDF-TrFE piezoelectric flow sensor integrated in a tube was able to measure flow rates with the sensitivity of 0.02 mVpp per 20 mL/min in the range of portal vein flow rate. It also showed excellent operational stability for 120 minutes with a standard deviation of 0.084 mVpp. The tube can be used as a clinical tool for assessing the need for portal vein inflow modification during liver transplantation.
  • Keywords
    blood flow measurement; blood vessels; flow sensors; intelligent sensors; liver; piezoelectric devices; flow rate; integrated piezoelectric flow sensor; liver transplantation; operational stability; portal vein; sensing tube; smart tube; time 120 min; Biosensing Techniques; Blood Flow Velocity; Electromagnetic Fields; Equipment Design; Liver Transplantation; Polyvinyls; Portal Vein;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 2009. EMBC 2009. Annual International Conference of the IEEE
  • Conference_Location
    Minneapolis, MN
  • ISSN
    1557-170X
  • Print_ISBN
    978-1-4244-3296-7
  • Electronic_ISBN
    1557-170X
  • Type

    conf

  • DOI
    10.1109/IEMBS.2009.5333823
  • Filename
    5333823