• DocumentCode
    2401260
  • Title

    Pulse transit time-based blood pressure estimation using hilbert-huang transform

  • Author

    Zhang, Qiao ; Shi, Yang ; Teng, Daniel ; Dinh, Anh ; Ko, Seok-Bum ; Chen, Li ; Basran, Jenny ; Dal Bello-Haas, Vanina ; Choi, Younhee

  • Author_Institution
    Div. of Biomed. Eng., Univ. of Saskatchewan, Saskatoon, SK, Canada
  • fYear
    2009
  • fDate
    3-6 Sept. 2009
  • Firstpage
    1785
  • Lastpage
    1788
  • Abstract
    The pulse transit time (PTT) based method has been suggested as a continuous, cuffless and non-invasive approach to estimate blood pressure. It is of paramount importance to accurately determine the pulse transit time from the measured electrocardiogram (ECG) and photoplethysmo-gram (PPG) signals. We apply the celebrated Hilbert-Huang Transform (HHT) to process both the ECG and PPG signals, and improve the accuracy of the PTT estimation. Further, the blood pressure variation is obtained by using a well-established formula reflecting the relationship between the blood pressure and the estimated PTT. Simulation results are provided to illustrate the effectiveness of the proposed method.
  • Keywords
    Hilbert transforms; blood pressure measurement; electrocardiography; plethysmography; Hilbert-Huang transform; blood pressure estimation; electrocardiography; photoplethysmography; pulse transit time; Blood Pressure; Electrocardiography;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 2009. EMBC 2009. Annual International Conference of the IEEE
  • Conference_Location
    Minneapolis, MN
  • ISSN
    1557-170X
  • Print_ISBN
    978-1-4244-3296-7
  • Electronic_ISBN
    1557-170X
  • Type

    conf

  • DOI
    10.1109/IEMBS.2009.5334008
  • Filename
    5334008