• DocumentCode
    2406115
  • Title

    Millimetre wave primary noise standards in waveguide bands 18–170 GHz

  • Author

    Vizard, D.R. ; Foster, P.R. ; Lunn, B. ; Cherry, S.M.

  • Author_Institution
    VivaTech Ltd., Nice
  • fYear
    2007
  • fDate
    9-12 Oct. 2007
  • Firstpage
    352
  • Lastpage
    355
  • Abstract
    Accurate measurement of millimetre wave systems in terms of their noise performance is dependent on the evolution of appropriate and convenient standards. The popular use of noise diodes as a transfer standard have limitations in terms of their frequency coverage and absolute accuracy. Consequently significant errors arise in the calibration of sensitive radiometers and other low noise components or instruments. A convenient laboratory cryogenic primary standard is described which allows improved accuracy whilst extending the prior art to frequencies as high as 170 GHz.
  • Keywords
    calibration; cryogenics; measurement errors; measurement standards; microwave measurement; millimetre wave measurement; noise measurement; radiometry; calibration; frequency 18 GHz to 170 GHz; laboratory cryogenic primary standard; measurement errors; millimetre wave primary noise standards; millimetre wave systems; noise diodes; radiometers; waveguide band; Art; Calibration; Cryogenics; Diodes; Frequency; Instruments; Laboratories; Measurement standards; Noise measurement; Radiometers;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Microwave Conference, 2007. European
  • Conference_Location
    Munich
  • Print_ISBN
    978-2-87487-001-9
  • Type

    conf

  • DOI
    10.1109/EUMC.2007.4405199
  • Filename
    4405199