DocumentCode :
2411766
Title :
Medical safety of TASER conducted energy weapon in a hybrid 3-point deployment mode
Author :
Panescu, Dorin ; Kroll, Mark W. ; Stratbucker, Robert A.
Author_Institution :
NewCardio, Inc., Santa Clara, CA, USA
fYear :
2009
fDate :
3-6 Sept. 2009
Firstpage :
3191
Lastpage :
3194
Abstract :
Introduction: TASER conducted energy weapons (CEW) deliver electrical pulses that can temporarily incapacitate subjects. The goal of this paper is to analyze the distribution of TASER CEW currents in the heart and surrounding organs and to understand theoretical chances of triggering cardiac arrhythmias, of capturing the vagus and phrenic nerves and producing electroporation of skeletal muscle structures. The CEW operates in either probe mode or drive-stun (direct contact) mode. There is also a hybrid mode in which current is passed from a single probe to either or both of 2 drive-stun electrodes on the weapon, presumed to be in direct contact with the skin. Methods and Results: The models analyzed herein describe strength-duration thresholds for myocyte excitation and ventricular fibrillation (VF) induction. Finite element modeling (FEM) was used to approximate current density in the heart for worst-case TASER electrode placement. The FEMs theoretically estimated that maximum TASER CEW current densities in the heart and in neighboring organs are at safe levels. A 3-point deployment mode was compared to probe-mode deployment. The margins of safety for the 3-point deployment were estimated to be as high as or higher than for the probe-mode deployment. Conclusion: Numerical modeling estimated that TASER CEWs were expected to be safe when deployed in 3-point mode. In drive-stun, probe-mode or 3-point deployments, the CEWs had high theoretically approximated safety margins for cardiac capture, VF, phrenic or vagus nerve capture and skeletal muscle damage by electroporation.
Keywords :
bioelectric phenomena; biological effects of fields; biological organs; bone; cardiology; cellular biophysics; current density; finite element analysis; health and safety; muscle; neurophysiology; skin; weapons; FEM; TASER conducted energy weapon; TASER electrode placement; cardiac arrhythmia; cardiac capture; current density; drive-stun electrodes; electrical pulses; electroporation; finite element modeling; hybrid three-point deployment mode; medical safety; myocyte excitation; phrenic nerve; skeletal muscle structures; vagus nerve; ventricular fibrillation induction; Arrhythmia; TASER; cardiac; fibrillation; modeling; Arrhythmias, Cardiac; Electric Stimulation; Electroporation; Electroshock; Equipment Design; Finite Element Analysis; Humans; Models, Theoretical; Muscle Contraction; Muscle, Skeletal; Reproducibility of Results; Skin; Software; Weapons;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society, 2009. EMBC 2009. Annual International Conference of the IEEE
Conference_Location :
Minneapolis, MN
ISSN :
1557-170X
Print_ISBN :
978-1-4244-3296-7
Electronic_ISBN :
1557-170X
Type :
conf
DOI :
10.1109/IEMBS.2009.5334538
Filename :
5334538
Link To Document :