DocumentCode :
2413130
Title :
Measurements of cerebral arterial oxygen saturation using a fiber-optic pulse oximeter
Author :
Phillips, Justin P. ; Langford, Richard M. ; Chang, Serene H. ; Maney, Kishore ; Kyriacou, Panayiotis A. ; Jones, Deric P.
Author_Institution :
Sch. of Eng. & Math. Sci., City Univ., London, UK
fYear :
2009
fDate :
3-6 Sept. 2009
Firstpage :
1509
Lastpage :
1512
Abstract :
This pilot investigation was undertaken to assess the performance of a novel fiber-optic cerebral pulse oximetry system. A fiber-optic probe designed to pass through the lumen of a cranial bolt of the type used to make intracranial pressure measurements was used to obtain optical reflectance signals directly from the brain tissue. Preliminary results from seven patients measured in the operating theatre and ITU are presented. Estimations of cerebral arterial oxygen saturation derived from a frequency domain-based algorithm are compared with pulse oximetry (SpO2) and hemoximeter (SaO2) blood samples. The mean (plusmnSD) difference between cerebral oxygen saturation (ScaO2) and finger SpO2 (in saturation units) was -7.47(plusmn3 .4)%. The mean (plusmnSD) difference between ScaO2 and blood SaO2 was -7.37(plusmn2.8)%.
Keywords :
bio-optics; brain; fibre optic sensors; medical signal processing; oximetry; oxygen; O2; blood; brain tissue; cerebral arterial oxygen saturation; fiber-optic pulse oximeter; frequency domain-based algorithm; hemoximeter; optical reflectance signals; Blood; Cranial pressure; Fasteners; Frequency estimation; Optical design; Optical pulses; Pressure measurement; Probes; Pulse measurements; Signal design;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society, 2009. EMBC 2009. Annual International Conference of the IEEE
Conference_Location :
Minneapolis, MN
ISSN :
1557-170X
Print_ISBN :
978-1-4244-3296-7
Electronic_ISBN :
1557-170X
Type :
conf
DOI :
10.1109/IEMBS.2009.5334604
Filename :
5334604
Link To Document :
بازگشت