• DocumentCode
    242249
  • Title

    A low power low niose AFE with continuous time self-calibration filters for ECG monitor and heart beat detection

  • Author

    Jiali Hou ; Weibo Hu ; Ke Lin ; Bo Wang ; Chaoxun Wang ; Mo Wang ; Kai Xu ; Xin´an Wang

  • Author_Institution
    Shenzhen Grad. Sch., Sch. of ECE, Peking Univ., Shenzhen, China
  • fYear
    2014
  • fDate
    28-31 Oct. 2014
  • Firstpage
    1
  • Lastpage
    3
  • Abstract
    A low-power low-noise analog front end (AFE) circuit is presented for ECG monitor and QRS feature extraction. For the purpose of low power and small chip area, a 1st-order high pass filter with tunable pseudo-resistor and small capacitor is proposed, and a low pass Gm-C filter using capacitance scaler technique is adopted. A novel calibration circuit is proposed to calibrate the cut-off frequency of the low-pass filter which is sensitive to PVT. The AFE consumes 10μW and input-referred integrated noise in the signal band is 6μVrms.
  • Keywords
    analogue circuits; electrocardiography; high-pass filters; low-pass filters; low-power electronics; 1st-order high pass filter; ECG monitor; PVT; QRS feature extraction; calibration circuit; capacitance scaler; continuous time self-calibration filters; heart beat detection; input-referred integrated noise; low pass filter; low power low niose analog front end; power 10 muW; Abstracts; Biomedical monitoring; Calibration; Feature extraction; Gain measurement; Logic gates; Monitoring;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Solid-State and Integrated Circuit Technology (ICSICT), 2014 12th IEEE International Conference on
  • Conference_Location
    Guilin
  • Print_ISBN
    978-1-4799-3296-2
  • Type

    conf

  • DOI
    10.1109/ICSICT.2014.7021551
  • Filename
    7021551