• DocumentCode
    2424128
  • Title

    Design and performance evaluation of a high resolution small animal positron tomograph

  • Author

    Pavlopoulos, S. ; Tzanakos, George

  • Author_Institution
    Dept. of Biomed. Eng., Rutgers Univ., Piscataway, NJ, USA
  • fYear
    1993
  • fDate
    31 Oct-6 Nov 1993
  • Firstpage
    1697
  • Abstract
    The authors have used Monte Carlo simulation methods to design and evaluate a small animal positron tomograph of inner diameter 34.37 cm, field of view diameter of 20 cm, and axial field of view of 6 cm. The tomograph consists of 16 BGO blocks, each of which is partially segmented to 18×16 crystals of dimension 5.35 mm×3.30 mm×30 mm and is read out by a position sensitive photomultiplier. The rate performance as well as the spatial resolution have been studied and compared with other tomographs
  • Keywords
    Monte Carlo methods; biological techniques; performance evaluation; positron emission tomography; 6 to 34.37 cm; BGO blocks; Monte Carlo simulation methods; PET performance evaluation; PET system design; high resolution small animal positron tomograph; partially segmented blocks; physiological research; rate performance; spatial resolution; Animals; Cathodes; Detectors; Energy resolution; Imaging phantoms; Photomultipliers; Photonic crystals; Positron emission tomography; Spatial resolution; Timing;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium and Medical Imaging Conference, 1993., 1993 IEEE Conference Record.
  • Conference_Location
    San Francisco, CA
  • Print_ISBN
    0-7803-1487-5
  • Type

    conf

  • DOI
    10.1109/NSSMIC.1993.373580
  • Filename
    373580