    A PET scatter correction using simultaneous acquisitions with low and high lower energy thresholds

  • Author

    Bendriem, B. ; Bossen, R. Tre ; Frouin, V. ; Syrota, A.

  • Author_Institution
    Service Hospitalier F. Joliot CEA, Orsay, France
  • fYear
    1993
  • fDate
    31 Oct-6 Nov 1993
  • Firstpage
    1779
  • Abstract
    The authors propose a new method for scatter correction using two sinograms acquired simultaneously with a low and a high lower energy threshold (LLD). The high LLD is chosen such that the corresponding sinogram contains no or few scatter counts. The information in each data set are combined to produce a sinogram corrected for scatter. The method is described in detail and evaluated on various phantom measurements
  • Keywords
    medical image processing; positron emission tomography; PET scatter correction; high lower energy threshold; low energy threshold; medical diagnostic imaging; nuclear medicine; phantom measurements; scatter counts; simultaneous acquisitions; sinogram; Cameras; Computed tomography; Deconvolution; Electromagnetic scattering; Energy resolution; Imaging phantoms; Particle scattering; Photonic crystals; Positron emission tomography; Single photon emission computed tomography;