• DocumentCode
    2424740
  • Title

    Design and performance evaluation of a new high resolution detector array module for PET

  • Author

    Tzanakos, George ; Pavlopoulos, Sotiris

  • Author_Institution
    Dept. of Phys., Athens Univ., Greece
  • fYear
    1993
  • fDate
    31 Oct-6 Nov 1993
  • Firstpage
    1842
  • Abstract
    The authors have designed a 2-D detector-array module for high resolution PET. The module consists of a 16×18 matrix of crystals cut from a single BGO block coupled to a 3"×3" square envelope position-sensitive PMT. The performance of the detector module was studied using simulation techniques and the design was optimized to provide high resolution, good energy and timing characteristics, and ease in the construction. The energy resolution of the module at the photopeak was estimated to 27%, whereas the timing resolution of a coincident pair of detector modules was found to be 6.5 ns FWHM
  • Keywords
    biomedical equipment; design engineering; performance evaluation; positron emission tomography; 6.5 ns; BGO block; PET detector; coincident pair resolution; crystal matrix; energy resolution; high resolution detector array module; medical diagnostic imaging; medical instrumentation; nuclear medicine; optimized design; photopeak; simulation techniques; timing resolution; Anodes; Costs; Crystals; Energy resolution; Photomultipliers; Position sensitive particle detectors; Positron emission tomography; Sensor arrays; Spatial resolution; Timing;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium and Medical Imaging Conference, 1993., 1993 IEEE Conference Record.
  • Conference_Location
    San Francisco, CA
  • Print_ISBN
    0-7803-1487-5
  • Type

    conf

  • DOI
    10.1109/NSSMIC.1993.373612
  • Filename
    373612