Design and performance evaluation of a new high resolution detector array module for PET
Author
Tzanakos, George ; Pavlopoulos, Sotiris
Author_Institution
Dept. of Phys., Athens Univ., Greece
fYear
1993
fDate
31 Oct-6 Nov 1993
Firstpage
1842
Abstract
The authors have designed a 2-D detector-array module for high resolution PET. The module consists of a 16×18 matrix of crystals cut from a single BGO block coupled to a 3"×3" square envelope position-sensitive PMT. The performance of the detector module was studied using simulation techniques and the design was optimized to provide high resolution, good energy and timing characteristics, and ease in the construction. The energy resolution of the module at the photopeak was estimated to 27%, whereas the timing resolution of a coincident pair of detector modules was found to be 6.5 ns FWHM
Keywords
biomedical equipment; design engineering; performance evaluation; positron emission tomography; 6.5 ns; BGO block; PET detector; coincident pair resolution; crystal matrix; energy resolution; high resolution detector array module; medical diagnostic imaging; medical instrumentation; nuclear medicine; optimized design; photopeak; simulation techniques; timing resolution; Anodes; Costs; Crystals; Energy resolution; Photomultipliers; Position sensitive particle detectors; Positron emission tomography; Sensor arrays; Spatial resolution; Timing;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium and Medical Imaging Conference, 1993., 1993 IEEE Conference Record.
Conference_Location
San Francisco, CA
Print_ISBN
0-7803-1487-5
Type
conf
DOI
10.1109/NSSMIC.1993.373612
Filename
373612
Link To Document