    Development of equipment to provide monochromatic light over wavelengths from 367–850 nm for measurement of pupillary behavior

  • Author

    Canul-Gómez, G.D. ; Suaste-Gómez, E.

  • Author_Institution
    Electr. Eng. Dept., CINVESTAV Mexico, Mexico City, Mexico
  • fYear
    2009
  • fDate
    16-20 March 2009
  • Firstpage
    8
  • Lastpage
    8
  • Abstract
    The importance of analyzing the behavior of biological tissue when exposed to colored light has opened new insights into the medical field for diagnosis of various pathologies. This type of study is achieved with a computer controlled monochromator over the range 367-850 nm; near uv to near ir. The equipment consists of coupling convex lenses that focus a light source of intensity 7500 lumens to the slot of the monochromator. The monochromator has a resolution of 1 nm and a bandwidth of 10 nm. This equipment is being used as a slit lamp to determine changes in the pupillary behavior with respect to wavelength.
  • Keywords
    bio-optics; biological tissues; biomedical imaging; diseases; eye; a monochromator; biological tissue; colored light; coupling convex lenses; monochromatic light; pathologies; patient diagnosis; pupillary; wavelength 367 nm to 850 nm; Bandwidth; Biological tissues; Biology computing; Lenses; Light sources; Medical diagnostic imaging; Optical computing; Optical coupling; Pathology; Wavelength measurement;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Health Care Exchanges, 2009. PAHCE 2009. Pan American
  • Conference_Location
    Mexico City
  • Print_ISBN
    978-1-4244-3668-2
  • Electronic_ISBN
    978-1-4244-3669-9
  • Type

    conf

  • DOI
    10.1109/PAHCE.2009.5158351
  • Filename
    5158351