Abstract
A method to determine the step number of a Josephson array by using two combinations of frequency and step number is studied and uncertainty analysis is given. The method can correctly determine step number by using a voltmeter with accuracy of two orders of magnitude worse than that required in conventional method.
Keywords
Josephson effect; measurement standards; superconducting junction devices; voltage measurement; voltmeters; DC voltage measurement; Josephson junction array voltage standard; step number; uncertainty analysis; voltmeter; Equations; Frequency; Josephson junctions; Laboratories; Measurement standards; Millimeter wave measurements; Uncertainty; Virtual manufacturing; Voltage measurement; Voltmeters;