DocumentCode :
2443311
Title :
A CMOS-based phototransistor for high-sensitivity biochemical detection using absorption photometry
Author :
Chang, Yu-Wei ; Tai, Yu-Ting ; Huang, Yang-Tung ; Yang, Yuh-Shyong
Author_Institution :
Inst. of Electron., Nat. Chiao Tung Univ., Hsinchu
fYear :
2008
fDate :
Nov. 30 2008-Dec. 3 2008
Firstpage :
82
Lastpage :
85
Abstract :
This paper proposes a high-sensitivity biosensing system on the basis of absorption photometry and a phototransistor. Manufactured in a standard 0.35-mum CMOS technology, the phototransistor exhibits a responsivity higher than 1000 A/W for 650-nm light. The TMB/H2O2/HRP method is chosen as a useful platform to determine the amount of various interesting targets. For the biochemical detection, a sample volume of only 10 mul is required to be dropped on the slide above the phototransistor. Experimental results demonstrate that a high sensitivity of 2.5 muA/pM is achieved, and the minimum HRP concentration successfully detected is 2.7 pM. This detection limit is three orders of magnitude better than that of a reported silicon biosensor.
Keywords :
biomedical transducers; biosensors; photometry; phototransistors; CMOS technology; CMOS-based phototransistor; absorption photometry; biomedical transducers; high-sensitivity biochemical detection; high-sensitivity biosensing system; medical diagnosis; Biomedical transducers; medical diagnosis; photodetectors; phototransistors;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Sensing Technology, 2008. ICST 2008. 3rd International Conference on
Conference_Location :
Tainan
Print_ISBN :
978-1-4244-2176-3
Electronic_ISBN :
978-1-4244-2177-0
Type :
conf
DOI :
10.1109/ICSENST.2008.4757078
Filename :
4757078
Link To Document :
بازگشت