Abstract
During hyperthermia therapy for cancer treatment magnetic fluid is injected into the tumor. Hyperthermia therapy exploits the fact that cancer cells are more sensitive to heat than healthy cells. Generally all parameters except the magnetic fluid weight density are known in the specific heat equation which governs the heat given in hyperthermia therapy to destroy cancer cells. Hence, accurate estimation of magnetic fluid weight density inside the body is critical for successful treatment. A unique needle-type GMR sensor is fabricated for inserting into the body in a low-invasive way. This paper presents analytical analysis and experimental results obtained by estimating magnetic fluid weight density in large cavities by the fabricated needle-type GMR sensor.
Keywords
cancer; density measurement; giant magnetoresistance; hyperthermia; magnetic field effects; magnetic fluids; magnetic permeability; magnetic variables measurement; patient treatment; demagnetizing factor; hyperthermia therapy cancer treatment; large cavity; magnetic fluid weight density estimation; needle-type GMR sensor; relative permeability; Cancer; Hyperthermia; Magnetic flux; Magnetic liquids; Magnetic sensors; Magnetic separation; Medical treatment; Nanoparticles; Neoplasms; Tumors; GMR sensor; demagnetizing factor; magnetic nanoparticles; relative permeability; tumor;
fLanguage
English
Publisher
ieee
Conference_Titel
Sensing Technology, 2008. ICST 2008. 3rd International Conference on
Conference_Location
Tainan
Print_ISBN
978-1-4244-2176-3
Electronic_ISBN
978-1-4244-2177-0
Type
conf
DOI
10.1109/ICSENST.2008.4757184
Filename
4757184
Link To Document