DocumentCode :
2449799
Title :
The power supply for a medical synchrotron Beam Chopper System
Author :
Dallago, E. ; Venchi, G. ; Rossi, S. ; Pullia, M. ; Fowler, T. ; Nielsen, U.
Author_Institution :
Dept. of Electr. Eng., Univ. of Pavia, Pavia
fYear :
2008
fDate :
10-13 Nov. 2008
Firstpage :
1016
Lastpage :
1020
Abstract :
In this paper the power supply for the Beam Chopper System of a medical synchrotron is presented. The four, series connected beam chopper magnets, which are placed on the extraction line, are used for blanking the beam both during routine operation and for emergency. To let the beam pass the magnets need to be energized with a current, proportional to the beam energy, whose maximum value is 650 A. Rise and fall time shall be less than 260 mus, while the flat-top can be maintained for an arbitrary long time. The power supply is composed of an high voltage part, which is in charge of the rising and falling current ramp, and of a low voltage part, which is in turn responsible for the current flat top. The paper presents details about the design and experimental results measured on the actual device.
Keywords :
beam choppers; biomedical equipment; power supplies to apparatus; synchrotron radiation; beam chopper system; beam pass; current 650 A; current flat top; medical synchrotron; power supply; routine operation; time 260 mus; voltage part; Acceleration; Blanking; Choppers; Circuit topology; Instruments; Low voltage; Magnets; Power supplies; Protons; Synchrotrons;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Industrial Electronics, 2008. IECON 2008. 34th Annual Conference of IEEE
Conference_Location :
Orlando, FL
ISSN :
1553-572X
Print_ISBN :
978-1-4244-1767-4
Electronic_ISBN :
1553-572X
Type :
conf
DOI :
10.1109/IECON.2008.4758093
Filename :
4758093
Link To Document :
بازگشت