DocumentCode
2452484
Title
Comparison of rf power meters using two air-line impedance standards
Author
Stumper, U.
Author_Institution
Physikalisch Tech. Bundesanstalt, Braunschweig, Germany
fYear
1996
fDate
17-21 June 1996
Firstpage
382
Lastpage
383
Abstract
A method to calibrate a coaxial RF power meter by comparison with a standard power meter is described, where the ratio of the mismatch factors is determined using two calculable coaxial air-line impedance standards. First experimental results of measurements performed in the coaxial 7 mm/N technique at frequencies from 2 GHz to 18 GHz are presented.
Keywords
calibration; measurement standards; microwave measurement; power measurement; power meters; 2 to 18 GHz; air-line impedance standards; coaxial RF power meter; coaxial air-line impedance standards; mismatch factors; rf power meters; standard power meter; Calibration; Coaxial components; Impedance; Measurement standards; Performance evaluation; Phase measurement; Power measurement; Radio frequency; Reflection; Wavelength measurement;
fLanguage
English
Publisher
ieee
Conference_Titel
Precision Electromagnetic Measurements Digest, 1996 Conference on
Conference_Location
Braunschweig, Germany
Print_ISBN
0-7803-3376-4
Type
conf
DOI
10.1109/CPEM.1996.547125
Filename
547125
Link To Document